• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ UD 27IN 3-0 S/A SH-1; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ UD 27IN 3-0 S/A SH-1; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Model Number VCP219H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device not returned.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Sample analysis: visual analysis of the returned sample determined that one empty winding former, a paper lid and a needle product code vcp219h were received to ethicon for evaluation.The product code is single-armed and on the needle, the swage and attachment area were noted to be as expected, marks by the surgical instrument were noted on body needle and no suture remnant could be observed into the barrel hole.The suture was not received for evaluation to determine the assignable cause.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Trade name: irgacare®, active ingredient(s): triclosan, dosage form: suture/solid/parenteral, strength: = 2360 g/m.
 
Event or Problem Description
It was reported that a patient underwent a thyroidectomy surgery on (b)(6) 2022 and suture was used.During the procedure, the reported suture detached from the swaged end of the needle during usage.No adverse patient consequences were reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 27IN 3-0 S/A SH-1
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
avenida de las torres 7125
col salvacar
cuidad juarez
MX  
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key15015417
Report Number2210968-2022-05471
Device Sequence Number15139411
Product Code GAM
UDI-Device Identifier10705031051720
UDI-Public10705031051720
Combination Product (Y/N)Y
Initial Reporter CountryHK
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVCP219H
Device Catalogue NumberVCP219H
Device Lot NumberSBMCAH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Initial Report FDA Received Date07/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-