• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. CPS SHORT ANCHOR PLUG 16MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. CPS SHORT ANCHOR PLUG 16MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Fall (1848); Pain (1994)
Date of Event 06/24/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision surgery approximately 1 month post implantation s the patient fell and broke the anchor plug.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Foreign: (b)(6).No devices or photographs were received; therefore, the condition of the components is unknown.The device history record was reviewed and no discrepancies relevant to the reported event were found.It was stated that the patient fell, however it is unknown if the implants caused/contributed to the fall.Therefore, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that initial right oss operation was performed.Approximately 13 months post implantation, the anchor plug broke after the patient fell.The patient was revised.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: a2; a3; b4; b5; b7; g3; g6; h1; h2.An investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6; h11 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Journal article provided confirms the patient's anchor plug was fractured.X-rays provided were not sent for review since the journal article confirms the anchor plug fracture.Patient fell at 13 months however the cause of fall is unknown.A definitive root cause cannot be determined.This complaint is confirmed via journal article notes provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CPS SHORT ANCHOR PLUG 16MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15016257
Report Number0001825034-2022-01641
Device Sequence Number11366199
Product Code KWA
UDI-Device Identifier00880304461673
UDI-Public(01)00880304461673(17)271127(10)671350
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K062998
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number178558
Device Lot Number671350
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer07/30/2022
08/28/2025
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date08/21/2025
09/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/27/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age36 YR
Patient SexMale
-
-