• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PRODIGY MRI IPG, 16CH, RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC) PRODIGY MRI IPG, 16CH, RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3772
Device Problem Wireless Communication Problem (3283)
Patient Problem Inadequate Pain Relief (2388)
Event Date 10/15/2021
Event Type  Injury  
Event Description
It was reported that patient unable to successfully recharge ipg and subsequently lost stimulation.Replacement of charger did not resolve the issue.As such, the ipg became inoperable and was unable to communicate with all external devices.In turn, the patient underwent surgical intervention wherein the ipg was explanted and replaced.
 
Manufacturer Narrative
A review of documentation supplied with the implant states that the implant must be charged every 30 to 90 days to avoid battery depletion.Based on the information received, the cause of the reported incident is consistent with user error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODIGY MRI IPG, 16CH, RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15016931
MDR Text Key295904851
Report Number3006705815-2022-15541
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017093
UDI-Public05415067017093
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/04/2019
Device Model Number3772
Device Catalogue Number3772
Device Lot NumberA000040052
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/27/2022
Initial Date FDA Received07/14/2022
Supplement Dates Manufacturer Received01/29/2023
Supplement Dates FDA Received01/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/04/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS LEAD(1); SCS LEAD(1); SCS LEAD(1); SCS LEAD(1)
Patient Outcome(s) Other;
Patient Age35 YR
Patient SexFemale
Patient Weight68 KG
-
-