Model Number 45030
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the catheter stuck on wire.An angiojet solent dista catheter was used for thrombectomy procedure.However, during the procedure, the catheter became stuck on wire and could not advance.The procedure was completed via alternative method or device.No patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of the solent dista thrombectomy with an unknown guidewire stuck inside the device.The pump, effluent/supply line, shaft, tip, piston, and spike line were visually inspected.Visual examination revealed a flat spot to the shaft 64.5cm from the strain relief.Microscopic examination revealed no additional damages.Inspection of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that the catheter stuck on wire.An angiojet solent dista catheter was used for thrombectomy procedure.However, during the procedure, the catheter became stuck on wire and could not advance.The procedure was completed via alternative method or device.No patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOJET SOLENT DISTA
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15017276
Report Number2124215-2022-25018
Device Sequence Number20373988
Product Code DXE
UDI-Device Identifier08714729889731
UDI-Public08714729889731
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K113363
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number45030
Device Catalogue Number45030
Device Lot Number0028535798
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer08/04/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date08/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/31/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-