PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Codes Failure to Analyze Signal (1539); Device Sensing Problem (2917); Output Problem (3005)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker evaluated the customer¿s device and verified the reported issue.Stryker replaced the therapy pcb assembly and then observed proper device operation through functional and performance testing.The device was returned to the customer for use.
 
Event or Problem Description
The customer contacted stryker to report that their service indicator was illuminated.In addition, their device had logged an event code in the memory which indicates that the aed function and paddles ecg of the device may be inoperable.In this state the device may not be able to analyze a patient's rhythm and provide defibrillation therapy while in aed mode, if necessary.There was no patient use associated with the reported event.
 
Additional Manufacturer Narrative
Stryker further evaluated the removed therapy pcb assembly and determined that the cause of the reported issue was due to capacitor, designator c3.C3 was electrically shorted.
 
Event or Problem Description
The customer contacted stryker to report that their service indicator was illuminated.In addition, their device had logged an event code in the memory which indicates that the aed function and paddles ecg of the device may be inoperable.In this state the device may not be able to analyze a patient's rhythm and provide defibrillation therapy while in aed mode, if necessary.There was no patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 15 MONITOR/DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key15017380
Report Number0003015876-2022-01459
Device Sequence Number14056991
Product Code MKJ
UDI-Device Identifier00883873979358
UDI-Public00883873979358
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-000061
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer12/27/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date12/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/29/2010
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-