Catalog Number INSC-7-230-S
Device Problems Separation Failure (2547); Failure to Fire (2610)
Patient Problem Gastrointestinal Hemorrhage (4476)
Event Type  Injury  
Event Description
During an unspecified procedure, the physician used a cook instinct endoscopic hemoclip.It was reported that the physician tried to deploy clip but it wouldn't deploy and the jaws would not open.They were unable to get the clip to release from the catheter or release from the tissue.The nurses and doctor jiggled the catheter but the clip still wouldn't deploy.The clip eventually deployed but the patient presented for re-bleeding a few days later.An unintended section of the device did not remain inside the patient¿s body.Patient presented with rebleeding days later and required another endoscopy procedure.
 
Manufacturer Narrative
Pma/510(k): k212323.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The instructions for use state: "to permanently deploy clip, pull handle spool toward handle thumb ring until clip detaches.Note: if separation of clip is not immediate, gently move catheter back and forth or use other endoscopic maneuvers to separate catheter from clip." the instructions for use include the following precaution: "if clip deployment device is used with endoscope in a torqued or retroflexed position, clip deployment difficulties can occur." prior to distribution, all instinct endoscopic hemoclips are subjected to a visual and functional inspection to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSTINCT ENDOSCOPIC HEMOCLIP
Type of Device
PKL, LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key15017603
MDR Text Key295953564
Report Number1037905-2022-00389
Device Sequence Number1
Product Code PKL
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/14/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberINSC-7-230-S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/20/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
OLYMPUS SCOPE UNKNOWN MODEL
Patient Outcome(s) Required Intervention;
-
-