• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLSTENT ENDOPROSTHESIS; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLSTENT ENDOPROSTHESIS; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number 26280
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter facility name: (b)(6).
 
Event or Problem Description
It was reported that stent damage occurred.During unpacking of a 12x90/8fr uni plus halo 75cm wallstent endoprosthesis, it was noted that the stent struts were damaged.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
 
Additional Manufacturer Narrative
E1 - initial reporter facility name: (b)(6).Device evaluated by mfr: a 12x90/8fr uni plus halo 75cm was received for analysis.The device was received with the stent mounted in the correct position on the delivery system.The investigator was unable to deploy the stent as indicated in the instructions for use due to a severe kink to the stainless-steel shaft.To facilitate analysis of the stent the investigator dissected the shaft of the device and deployed the stent by grasping the tip and withdrawing the inner shaft.A visual examination of the stent found no issues or damage to the stent.A visual and tactile inspection found the shaft of the device to be severely kinked at more than one location.This type of damage is consistent with excessive force being applied to the device.A visual and tactile examination identified no issues with the tip of the device.
 
Event or Problem Description
It was reported that stent damage occurred.During unpacking of a 12x90/8fr uni plus halo 75cm wallstent endoprosthesis, it was noted that the stent struts were damaged.The procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLSTENT ENDOPROSTHESIS
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15018244
Report Number2124215-2022-24029
Device Sequence Number6041259
Product Code MAF
UDI-Device Identifier08714729203971
UDI-Public08714729203971
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152842
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/07/2022
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/18/2022
Device Model Number26280
Device Catalogue Number26280
Device Lot Number0026540323
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/24/2022
Supplement Date Received by Manufacturer08/16/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/18/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
Patient Weight69 KG
-
-