| Model Number |
406844 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During a paroxysmal supraventricular tachycardia procedure, there was a blood leak from the hemostasis valve.The introducer was replaced and the procedure was completed with no consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8f fast-cath introducer sheath was received for evaluation.The sheath passed pressure and aspiration leak testing with no anomalies observed.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported leak remains unknown.
|
| |
|