• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEVRO CORP. SENZA; NEVRO SENZA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEVRO CORP. SENZA; NEVRO SENZA Back to Search Results
Model Number LEAD1058-90B
Medical Device Problem Codes Insufficient Device Problem Information (3190); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 06/15/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Nevro is awaiting the return of the device.The manufacturing records were reviewed and no relevant nonconformities were found.Nevro submits this report in compliance with fda¿s medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Event or Problem Description
It was reported that a lead electrode was sheared off during the implant procedure.The physician made the determination to leave the electrode in the patient's body.The remaining portion of the lead was removed and a replacement lead was implanted successfully.The patient continues to use the device to find effective pain relief and there have been no reports of further complications regarding this event.
 
Additional Manufacturer Narrative
Nevro submits this report in compliance with fda's medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.H3 other text : the device was not returned.
 
Event or Problem Description
Multiple attempts were made to obtain the device back for evaluation, but the device was not returned.
 
Additional Manufacturer Narrative
The device was returned and analyzed.Visual inspection confirmed damage to the distal end of the lead.Based on the deformation seen, the lead failed from tensile overload below the most distal contact with no signs of cutting.The investigation observed from the fluoroscopy images that the failure occurred in the vicinity of the insertion needle tip.However, it cannot be concluded why the tip of the lead caught at that location, nor what the nature of the tensile force that caused the failure was.The manufacturing records were reviewed and the device was correctly manufactured by trained operators with no associated nonconformities.The exact root cause of the damage could not be determined.Nevro submits this report in compliance with fda's medical device reporting regulations under 21 cfr part 803.Nevro has complied with regulatory investigation requirements and is submitting all information that is reasonably known to us at this time.However, we may not have been able to confirm this information or complete the investigation within the timeframe for filing this report.We may have given no response or an incomplete response to certain questions because we do not currently have information available to provide a complete response.If we later obtain any required information that was not available at the time of this initial report, we will submit a supplemental report.This report is not an admission by anyone that the product described in this report was defective, that it malfunctioned, or that it caused or contributed to the event described in this report.We may conclude that the device had no defect, did not malfunction, or did not cause or contribute to a reportable event.Some of the items on this form include forced-choice terms used by the fda for reporting purposes that do not necessarily reflect nevro¿s conclusions about the causes or nature of the event.This statement should be included with any freedom of information act response.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENZA
Common Device Name
NEVRO SENZA
Manufacturer (Section D)
NEVRO CORP.
1800 bridge parkway
redwood city CA 94065
Manufacturer Contact
vis ayer
1800 bridge parkway
redwood city, CA 94065
6504333213
MDR Report Key15019472
Report Number3008514029-2022-00336
Device Sequence Number6800700
Product Code LGW
UDI-Device Identifier00813426020053
UDI-Public00813426020053
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P130022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberLEAD1058-90B
Device Catalogue NumberLEAD1058-90B
Device Lot Number94512952
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer06/15/2022
12/16/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date09/09/2022
12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age42 YR
Patient SexFemale
-
-