| Model Number |
N164 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent fluctuations in impedance measurements with no conclusive evidence of a product malfunction; please refer to the description for more information regarding the specific circumstances of this event.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system had been experiencing jumps in right ventricular (rv) lead pacing impedances from 800 ohms to 1600 ohms.The representative wanted to know if this jumps could relate to a lead issue, technical services (ts) was consulted and reviewed the case and noted there was no evidence of a fractured lead or an issue with lead integrity.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent fluctuations in impedance measurements with no conclusive evidence of a product malfunction; please refer to the description for more information regarding the specific circumstances of this event.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system had been experiencing jumps in right ventricular (rv) lead pacing impedances from 800 ohms to 1600 ohms.The representative wanted to know if this jumps could relate to a lead issue, technical services (ts) was consulted and reviewed the case and noted there was no evidence of a fractured lead or an issue with lead integrity.This rv lead remains in service.No adverse patient effects were reported.Additional information was received and this crt-d and rv lead were surgically abandoned and replaced.No additional adverse patient effects were reported.
|
| |
|