CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 SYS PED.WITH PRECHAMBER; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX438T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 06/15/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing site: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 (#fx438t) was implanted.According to the complainant, the shunt system was believed to be blocked.The patient underwent a revision procedure.The complaint device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 3 years.Gender: female.Same event as medwatch#3004721439-2022-00200.
 
Additional Manufacturer Narrative
Investigation: visual inspection: the following observations were made during the visual inspection: deposits/tissue residues on the third-party product (connector between ventricular catheter and peritoneal catheter).Deposits in the pediatric prechamber.Permeability test: the test showed that the progav 2.0 connected with the other parts is permeable.Computer control test: the computer controlled test showed the opening pressure of the progav 2.0, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 16.08 cmh2o.This is within the specified tolerance of 17 cmh2o ± 3 cmh2o.Adjustability test: the progav 2.0 was found to be adjustable to all pressure settings.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: in order to verify whether the investigated valve was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valve is finally opened with a special tool.After dismantling of the valve, deposits were found in progav 2.0.Results: based on our investigation results, we can determine deposits.The deposits had no effect to the technical properties at the time of the investigation.The detected proteins in the valve can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Event or Problem Description
Update patient information: age: 3 years, 6 months; weight: 9.4 kilograms (kg); height: 89 centimeters (cm).Same event as medwatch #3004721439-2022-00200.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 SYS PED.WITH PRECHAMBER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key15020903
Report Number3004721439-2022-00199
Device Sequence Number6781719
Product Code JXG
UDI-Device Identifier04041906137434
UDI-Public4041906137434
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2023
Device Model NumberFX438T
Device Catalogue NumberFX438T
Device Lot Number20039036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2022
Supplement Date Received by Manufacturer08/26/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/24/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age3 YR
Patient SexFemale
Patient Weight9 KG
-
-