Model Number 1806-4260S
Medical Device Problem Codes Break (1069); Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that 2 drills broke during operation.The broken-off drills were then pierced and removed contralaterally.The procedure was delayed by 15 min.
 
Additional Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Additional Manufacturer Narrative
Please note correction to d4 lot #.The reported condition could be confirmed as the drill bit is broken.The device inspection revealed the following: the visual inspection has shown that the drill bit is broken off at the cutting flutes, the fragment was returned for evaluation.The microscopic inspection shows that the breakage surface reveals typical characteristics of a brittle fracture i.E.Relatively smoother surface at the center than at the edges.Since the breakage was brittle (sudden) in nature, there is evidence of abrupt features especially the plateau on one side of the fracture face.There are nicks and dents at the flutes in the area of the breakage visible.These damages are an indication for an metallic contact, most likely with the nail during drilling.A hardness test according to rockwell on the returned item indicates the material being in accordance to the values in the material specification.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation, the root cause was attributed to be user related.The appearance of the fracture zone shows that a mechanical overload by applying excessive torsional force, most likely caused by an metallic contact with the nail, did lead to the breakage of the drill bit.If any further information is provided, the investigation report will be updated.
 
Event or Problem Description
It was reported that 2 drills broke during operation.The broken-off drills were then pierced and removed contralaterally.The procedure was delayed by 15 min.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL, AO, STERILE T2 FEMUR Ø4,2X340 MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15020943
Report Number0009610622-2022-00328
Device Sequence Number12462602
Product Code LXH
UDI-Device Identifier04546540196774
UDI-Public04546540196774
Combination Product (Y/N)N
Initial Reporter CountryAU
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1806-4260S
Device Catalogue Number18064260S
Device Lot NumberKU139137
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/18/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-