Model Number BF-1T60
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A user facility submitted a repair request to the olympus service center, for the oes bronchofiberscope for leaking at the insertion part tube.Upon inspection and testing of the customer returned device, the coating of the insertion guide serpentine tube was found peeled off.(1 to 2 mm or more).This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.There was no patient involvement, or user injury associated with this event.
 
Additional Manufacturer Narrative
The customer suspected device was returned for investigation.The olympus service center confirmed the reported event during inspection and testing.Upon evaluation of the returned device the following defects were found, water tightness not maintained due to holes in the insertion part serpentine tube, bending angle is insufficient due to the elongation of the angle wire, curved tube closer to the blade, insertion part serpentine crushed, scratches on the insertion part serpentine, coating of the insertion part serpentine tube peeled off, curved rubber adhesive missing, universal code crushed, scratches on the universal code, scratches on the operation part, scratches on the eyepiece, scratches on grip, angle lever scratched, scratches found on up/down plate, and light guide connector scratched.The faulty parts were replaced, and the device was returned to the user facility.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, a root cause could not be identified.However, it was presumed that the causes were stress during operation, external factors or handling.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES BRONCHOFIBERSCOPE
Common Device Name
BRONCHOFIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15021034
Report Number9610595-2022-00009
Device Sequence Number6857334
Product Code EOQ
UDI-Device Identifier04953170339264
UDI-Public04953170339264
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K023984
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBF-1T60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2022
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2022
Supplement Date Received by Manufacturer08/29/2022
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date09/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/14/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1