Model Number BG102-350
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the water flow in a glaucoma implant shunt was checked during the pre-operative preparation stage and the water did not flow or it was difficult to get through.The customer then replaced it with a new product.The patient is ok.No other information was provided.
 
Additional Manufacturer Narrative
Implant date: if implanted, give date: not applicable, as there was no patient contact with the device.Explant date: if explanted, give date: not applicable, as there was no patient contact with the device.Initial reporter name and address:: telephone number: (b)(6).The device is not returned; therefore, a failure analysis of the complaint device cannot be completed.A review of the device record and complaint trending for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Section d9.Device available for evaluation? yes returned to manufacturer on: jul.12, 2022 section h3.Device evaluated manufacturer? yes device evaluation: the product was received in the original packaging.The product was disinfected according to procedure.Product evaluation was performed under magnification.Additional testing was not performed.The results of the inspection shows that there is no problem with the tubing.Based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAERVELDT SHUNT
Common Device Name
IMPLANT, EYE VALVE
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15021906
Report Number3012236936-2022-01772
Device Sequence Number20547217
Product Code KYF
UDI-Device Identifier05050474513006
UDI-Public(01)05050474513006(17)220904
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K955455
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/04/2022
Device Model NumberBG102-350
Device Catalogue Number23030818
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer01/26/2023
Initial Report FDA Received Date07/14/2022
Supplement Report FDA Received Date01/30/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-