• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA VERSAFITCUP CC TRIO FLAT PE LINER Ø 32 / E; HIP SHELL LINER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA VERSAFITCUP CC TRIO FLAT PE LINER Ø 32 / E; HIP SHELL LINER Back to Search Results
Model Number 01.26.3244STT
Health Effect - Clinical Code Osteolysis (2377)
Date of Event 06/09/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on (b)(6) 2022: lot 146002: 45 items manufactured and released on 31-oct-2014.Expiration date: 2019-09-30.No anomalies found related to the problem.To date, all items of the same lot have been sold without any similar reported event during the period of review.Visual inspection: from the received pieces, it was possible to note in the liner a visible ovalization of the internal seat for the femoral head; the removal screw was still inside the rim, and a check with the femoral ball head was performed, confirming the ovalization of the inner liner.The femoral head shows some scratches and surface damages, potentially compatible with the presence in articulation of a small, hard object.The most relevant finding from the explants is an evident deformation of the internal liner; a possible root cause of the event can be polyethylene wear enhanced by a third body trapped in articulation, but it is difficult to determine it with certainty.
 
Event or Problem Description
At about 7 years after the primary, the patient came in reporting pain in the hip.Revision surgery was performed and it was discovered eccentric wear of the liner with subsequent osteolysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSAFITCUP CC TRIO FLAT PE LINER Ø 32 / E
Common Device Name
HIP SHELL LINER
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15025736
Report Number3005180920-2022-00529
Device Sequence Number12476867
Product Code LZO
UDI-Device Identifier07630030807589
UDI-Public07630030807589
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K103352
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Model Number01.26.3244STT
Device Catalogue Number01.26.3244STT
Device Lot Number146002
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2022
Initial Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/31/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-