| Model Number |
15 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A stryker service representative performed an initial evaluation of the customer¿s device and was able to verify the reported issue.Stryker continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to power on.In this state the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient involvement reported with the event.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device failed to power on.In this state the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient involvement reported with the event.
|
| |
|
Additional Manufacturer Narrative
|
|
The cause of the reported issue was isolated to the system pcb.However, the device could not be repaired as the replacement parts are no longer available.The device was returned to the customer unrepaired.The customer was informed of the device's state and provided information on obtaining a replacement device.
|
| |
|