Model Number 04.01.0208
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 06/21/2022
Type of Reportable Event Serious Injury
Event or Problem Description
At about 8 months after the primary surgery, the patient came in reporting pain due to the disassociation of the glenosphere from the baseplate.The surgeon revised the glenosphere, screw and liner and the surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on 23-jun-2022: lot 2003816a: (b)(4) manufactured and released on 20-jan-2021.Expiration date: 2026-01-14.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Lot 2003816: 11 items manufactured and released on 03-nov-2020.Expiration date: 2025-10-23.No anomalies found related to the problem.Additional device involved: batch review performed on 23-jun-2022: reverse shoulder system 04.01.0191 threaded glenoid baseplate ø24.5x30 (k171058) lot 1906800: (b)(4) items manufactured and released on 12-feb-2020.Expiration date: 2025-02-02.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Clinical evaluation performed by medacta medical affairs manager: revision 8 months after the primary rsa in a male patient due to the disassociation of the glenosphere from the baseplate.From the radiographic images, it is visible the glenophere dissociated from the baseplate.This event possibly happened due to the fact that one of the bone screws could not reach the final position and this prevented the glenosphere from finding perfect stability on the lateral surface on the taper spigot of baseplate, despite of central glenophere screw being properly tightened.The surgeon revised the glenosphere, screw and liner and the surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE SHOULDER SYSTEM LAT. GLENOSPHERE 39XØ24.5
Common Device Name
SHOULDER GLENOSPHERE
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15028118
Report Number3005180920-2022-00539
Device Sequence Number17695857
Product Code HSD
UDI-Device Identifier07630040728072
UDI-Public07630040728072
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K193175
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/14/2026
Device Model Number04.01.0208
Device Catalogue Number04.01.0208
Device Lot Number2003816A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2022
Initial Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexMale
Patient RaceWhite
-
-