• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CCWTT6
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Serious Injury
Event or Problem Description
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, lens was explanted due to toric markings being in the wrong place.Additional information has been requested and received stated that the marks for the toric are in the wrong place on the lens.There is a manufacturing error.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was not observed.Toric axis marks were within specifications.Iol (intra ocular lens) returned wrapped in a wipe in a clear zip lock bag.Solution is dried on both surfaces of the optic and haptics.Both haptics are intact.The optic is cracked/fractured split in two and scratched/marked-rejectable.The iol met specifications when dimensionally measured for edge thickness and plan view.The axis marks are within specification.No root cause identified as the reported complaint "toric markings in wrong place" was not observed, the returned iol met specification when dimensionally measured for edge thickness and plan view, the axis marks are within specification.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15028189
Report Number9612169-2022-00340
Device Sequence Number20422465
Product Code MJP
UDI-Device Identifier00380652458849
UDI-Public00380652458849
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/04/2025
Device Model NumberCCWTT6
Device Catalogue NumberCCWTT6.135
Device Lot Number25322681
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/07/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date07/15/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-