• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-55
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported during a routine check the cs300 intra-aortic balloon pump (iabp) the doppler probe was recognized to be broken and needed to be replaced.It is unknown if there was any patient involvement.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge fse replaced the doppler probe (0154-01-0005).The unit then passed all performance according to factory specifications.The unit was returned to customer and cleared for clinical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 220V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key15028887
MDR Text Key301773730
Report Number2249723-2022-01725
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567112541
UDI-Public10607567112541
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-55
Device Catalogue Number0998-00-3023-55
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/24/2010
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-