• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION REFURBISHED ENDOSCOPE 5.5M; LAPAROSCOPE, GENERAL Back to Search Results
Model Number MCENDO550R
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event or Problem Description
The user facility reported to terumo cardiovascular that the scope was cloudy.It is unknown when this event occurred, whether the product was changed out, or if there was any effect on the patient or results of the surgery.Due to the unknown information for this event, it is being reported.Terumo continues to attempt to gain more information regarding this event from the user facility.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information and device evaluation) ; h3 (device evaluated by manufacturer) ; h4 (device manufacture date); h6 (identification of evaluation codes 10, 3331, 3224, 3259, 19).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 3331 - analysis of production records.Investigation findings #1: 3224 - optical problem identified.Investigation findings #2: 3259 - improper physical structure.Investigation conclusions: 19 - cause traced to user.The returned unit was visually inspected for damage and looked through the endoscope directly to read some print matter.The internal components was then inspected using a monocular, and it was discovered that there was a crack on one of the internal lenses which resulted in a cloudy visual field.Improper handling caused the internal lens to break.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Event or Problem Description
New information received that states, the event occurred during vein harvesting, the product was changed out, no blood loss, and the surgery was completed successfully with no patient effect.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on jul 15, 2022.Upon further investigation of the reported event, the following information is new and/or changed: b5 (updated describe event or problem) d5 (added operator of device) e1 (added initial reporter) e2 (updated health professional) e3 (added occupation) g2 (added report source) g3 (date received by manufacturer) g6 (indication that this is a follow-up report) h2 (follow-up due to additional information) h6 (adverse event problem 2199, 4582, 11).A second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFURBISHED ENDOSCOPE 5.5M
Common Device Name
LAPAROSCOPE, GENERAL
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
jamie quinlan
125 blue ball road
elkton, MD 21921
8002837866
MDR Report Key15029783
Report Number1124841-2022-00131
Device Sequence Number14576242
Product Code GCJ
UDI-Device Identifier00699753018952
UDI-Public(01)00699753018952
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K092789
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMCENDO550R
Device Catalogue NumberN/A
Device Lot Number814754
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/30/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Supplement Date Received by Manufacturer07/19/2022
09/09/2022
Initial Report FDA Received Date07/15/2022
Supplement Report FDA Received Date08/04/2022
09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/27/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-