• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC 6 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC 6 SHOOTER SAEED MULTI-BAND LIGATOR; MND, LIGATOR Back to Search Results
Model Number G24654
Medical Device Problem Code Premature Activation (1484)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.A possible contributing factor to premature band deployment includes allowing the trigger cord to become lodged between the barrel and the distal end of the endoscope.This can restrict trigger cord movement and result in premature band deployment if enough tension is applied to pull the cord free.Premature band deployment can also occur if the handle is placed in the firing position before the endoscope is in place inside the patient or even while winding the trigger cord.The user is advised not to apply tension to the trigger cord while winding it on the handle to avoid premature deployment of bands.The instructions for use caution the user: ¿with handle in two-way position, slowly rotate handle clockwise to wind trigger cord onto handle spool until it is taut.Note: care must be taken to avoid deploying a band while winding trigger cord." rapid rotation of the ligator handle can contribute to premature band deployment.The instructions for use direct the user: ¿maintain suction and deploy band by rotating handle clockwise until band release is felt, indicating deployment." prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During a procedure for banding of varices in the stomach, the physician used a cook 6 shooter saeed multi-band ligator.It was reported that the device was loaded into the endoscope correctly but when the physician slowly turned the handle there was no click or resistance.The turning mechanism was smooth and suddenly all 6 bands closed in on the varix.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: the product said to be involved was returned in an open box from the lot number provided in the report.The label matches the product returned.Our laboratory evaluation of the product said to be involved confirmed the report of premature deployment based on the fact that no bands remained on the barrel when received.A functional test could not be performed on the device since no bands were remaining on the barrel.A visual examination of the device was performed.The trigger cord was examined and all twelve deployment beads were present.The beads were verified for correct location.The beads were examined using magnification and found to be correctly filled and had no evidence of excess flash.The length of the trigger cord was measured between the knots and found is within the established tolerance.The trigger cord was intact and not broken.An evaluation of the handle wheel movement was performed and the wheel functioned as intended.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the returned device confirmed the report due to no bands remaining on the barrel.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.A possible contributing factor to premature band deployment includes allowing the trigger cord to become lodged between the barrel and the distal end of the endoscope.This can restrict trigger cord movement and result in premature band deployment if enough tension is applied to pull the cord free.Premature band deployment can also occur if the handle is placed in the firing position before the endoscope is in place inside the patient or even while winding the trigger cord.The user is advised not to apply tension to the trigger cord while winding it on the handle to avoid premature deployment of bands.The instructions for use caution the user: ¿with handle in two-way position, slowly rotate handle clockwise to wind trigger cord onto handle spool until it is taut.Note: care must be taken to avoid deploying a band while winding trigger cord." rapid rotation of the ligator handle can contribute to premature band deployment.The instructions for use direct the user: ¿maintain suction and deploy band by rotating handle clockwise until band release is felt, indicating deployment." prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6 SHOOTER SAEED MULTI-BAND LIGATOR
Common Device Name
MND, LIGATOR
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
WILSON COOK MEDICAL INC.
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston-salem, NC 27105
8002454707
MDR Report Key15029823
Report Number1037905-2022-00393
Device Sequence Number11940656
Product Code MND
UDI-Device Identifier00827002246546
UDI-Public(01)00827002246546(17)220629(10)W4486449
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/29/2022
Device Model NumberG24654
Device Catalogue NumberMBL-6-OV
Device Lot NumberW4486449
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer07/18/2022
Initial Report FDA Received Date07/15/2022
Supplement Report FDA Received Date08/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FUJINON ENDOSCOPE UNKNOWN MODEL
Patient Age6 YR
Patient SexMale
Patient Weight20 KG
-
-