• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM; PUMP, INFUSION, INSULIN Back to Search Results
Model Number 18239
Device Problems Electrical /Electronic Property Problem (1198); Battery Problem (2885); Electrical Power Problem (2925)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2022
Event Type  malfunction  
Event Description
It was alleged that the patient's personal diabetes manager (pdm) battery exploded, causing the pdm's screen to burn.Complainant alleges that the pdm was in the patients pocket when the event occurred.
 
Manufacturer Narrative
The device has not been returned/received to date.If the device is received, a supplemental report will be submitted with the investigation results.It was reported that the personal diabetes manager (pdm) experienced an electrical failure that caused a burn to the pdm screen.Lot release records were reviewed and the product lot met all acceptance criteria.Specifically, a pod is paired to a pdm and put through simulated use testing including communicating with the pdm, deployment, delivering fluid, occlusion detection, and freedom from hazard alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD DASH¿ INSULIN MANAGEMENT SYSTEM
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
michael spears
100 nagog park
acton, MA 01720
9786007000
MDR Report Key15029983
MDR Text Key304340395
Report Number3004464228-2022-12072
Device Sequence Number1
Product Code LZG
UDI-Device Identifier10385082000009
UDI-Public(01)10385082000009(10)L000439
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192659
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/15/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number18239
Device Catalogue NumberUSA1-D001-MG-USA1
Was Device Available for Evaluation? No
Date Manufacturer Received06/22/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age30 YR
Patient SexMale
-
-