• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYMMETRY; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYMMETRY; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 23295
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that catheter entrapment on guidewire occurred.A 4.5-4/4t/90 symmetry balloon catheter was selected for use.However, after a non-boston scientific guidewire was inserted into the guidewire lumen of the balloon, a resistance was felt at 40cm from the handle side.It was noted that the balloon catheter was stuck with the guidewire.The procedure was completed with another of the same device.There were no patient complications nor injuries reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: a symmetry device was returned for analysis.This symmetry device is recommended for use with guidewire.The customer guidewire was not returned for analysis.During product analysis the investigator was able to load the device on to a boston scientific zipwire with no issues noted.A visual and tactile examination identified that the balloon was tightly folded and had not been subject to positive pressure.No issues were noted which may have potentially contributed to the complaint incident.A visual and tactile examination identified no issues with the shaft of the device.A visual and tactile examination identified no issues with tip of this device.
 
Event or Problem Description
It was reported that catheter entrapment on guidewire occurred.A 4.5-4/4t/90 symmetry balloon catheter was selected for use.However, after a non-boston scientific guidewire was inserted into the guidewire lumen of the balloon, a resistance was felt at 40cm from the handle side.It was noted that the balloon catheter was stuck with the guidewire.The procedure was completed with another of the same device.There were no patient complications nor injuries reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMMETRY
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15031030
Report Number2134265-2022-07323
Device Sequence Number19979309
Product Code LIT
UDI-Device Identifier08714729181859
UDI-Public08714729181859
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143193
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/09/2025
Device Model Number23295
Device Catalogue Number23295
Device Lot Number0028651041
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date07/15/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE USED: FUGA.; GUIDEWIRE USED: FUGA.
Patient SexUnknown
-
-