| Model Number |
PPH01 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/23/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a hemorrhoidectomy the instrument did not trigger.They finished the procedure with a new instrument.No patient harm nor significant delay reported.Procedure longo.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch#: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 08/15/2022.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that the pph01 device arrived with the knife damaged, with some unformed and protruding staples on the pockets, the breakaway washer was present and with an off-center cut.Also, the driver guide was noted to be damaged.No functional test was performed due to condition of the device.It appears possible that the anvil was pushed far enough off center to result in an off center cut of the breakaway washer and damage the knife by pressing it hard enough against the anvil.This situation normally occurs when the tissue is not evenly distributed in the device.However, it could not be determined what may have caused the anvil to become off center.It should be noted that ensuring that the tissue thickness is within the indicated range, and that it is evenly distributed in the device.Excess tissue on one side may result in unacceptable staple formation and can result in staple line leakage.Please reference the instruction for use for more information.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|