| Model Number |
712120 |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the gastrostomy feeding tube was removed due to a "hole" in the balloon.A test was performed and the leak was identified.The procedure was performed on (b)(6) 2022 and the material was removed on (b)(6) 2022.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) was reviewed, and no abnormal process conditions were present during the manufacturing of the product that could have led to the reported condition.The dhr review showed that all acceptance criteria inspections per established sampling levels were within acceptable limits during the production process.No sample was returned for evaluation.Three (3) photos were provided, and one shows the leak in the balloon.The complaint will be reopened if a sample is received.The reported condition is confirmed through the photo.This product is manufactured by a supplier.The root cause of this issue will be determined by the supplier and documented under a supplier corrective action report.This complaint will be used for tracking and trending purposes.
|
| |
|