| Model Number |
TX30V |
| Medical Device Problem Codes |
Difficult to Open or Close (2921); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.The lot/batch was not provided; therefore, the manufacturing record evaluation could not be performed.Additional information received: the device was used in an open right partial resection.The forceps was used to remove the tissue caught into the groove of the stapler, and the device was removed.No bleeding or damage on tissue occurred due to the issue.Patient condition 4 days after the operation so far patient is stable.The surgeon commented that the stapler clamped too much tissue.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be field as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a respiratory surgery, the tissue around the blood vessels got caught in a groove on the side of the staple, and the device could not be removed.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 08/03/2022.D4: batch # 5ey8j.Investigation summary: the product was returned to ethicon endo surgery for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the tx30v device arrived with no apparent damaged and with a reload loaded on the device.The reload was received fully fired.The device was tested for functionality with a test reload and it fired and formed all the staples as intended.The staple line was complete, and the staples meet the staple form release criteria.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Although no conclusion could be reached on the cause of the reported event, the instructions for use do contain the following caution: prior to opening the instrument and releasing the tissue, the edge of the reload or the side of the anvil may be used as a guide to transect tissue (not vessels) or to excise tissue which is protruding through the jaws.This aids in cutting at a proper distance from the staple line.Open the jaws by pushing the release button and releasing the grasp of the closing trigger.The triggers and jaws will fully open, releasing the tissue.Remove the instrument.The event could not be confirmed as the device performed as expected and during the functional testing, the device opened and closed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|