| Model Number |
68-700 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Device evaluation: analysis of the returned device identified: crease flat and white particles inner device leak test and microscopic analysis was performed which identified: device no leak and voids observed in the textured part of the valve (vv), it is out of specification according to qa 199.06 the fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: void on valve texture side assessed as a workmanship.No leak was observed in the device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.The device has been explanted and replaced.
|
| |
|