• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE, INCORPORATED NUVASIVE PULSE SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE, INCORPORATED NUVASIVE PULSE SYSTEM; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 1881500
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
During the confirmation scan of a navigated l4 to l5 anterior lumbar interbody fusion procedure the surgeon noted that the left and right screws placed on l4 were malplaced.The surgeon noted that the navigation was presenting 2mm proud and 3mm caudal to the actual placement.The screws were removed, the patient was re-scanned, and the screws were replaced correctly under navigation.
 
Additional Manufacturer Narrative
An on-site representative was able to confirm the reported event.Pre-case navigational accuracy checks passed.Post-case accuracy checks passed with no differences seen from the pre-case accuracy check.Additionally, the tracking arrays were re-registered and accuracy of instruments was checked four (4) times but no cause could be determined.Development engineering reviewed the system's log following the case including the re-registration accuracy tests but were unable to identify a cause.The system was tested by a service engineer in the field on 20apr2022 and passed all applicable testing.Based on the information obtained, the root cause of the reported event cannot be determined but may be a result of inaccurate tracking array registration at the beginning of the case due to hospital lighting factors.No additional investigation can be completed.Labeling review: ".Warnings, cautions and precautions: read all instructions and understand all warnings and cautions before using the system or accessories.Failure to do so may lead to serious medical consequences.Refer to the instructions for use accompanying other nuvasive products before use with the system to assure proper and safe use of these devices.If system data acquisition seems inaccurate or if the software application does not initiate or malfunctions during use, and recommended steps to restore the system are not successful, abort use of the system.The accuracy of data acquisition can be influenced by the rotation of the passive reflective array relative to the camera.Cameras should be placed such that a direct line of sight to the operative area is established." "intra-operative warnings:.Assess navigational accuracy repeatedly throughout a procedure when using a surgical navigation system.A) reconfirm accuracy by positioning the navigated instrument tip on an identifiable anatomical landmark and comparing the actual tip location to that displayed by the system.B) if the stereotaxic navigation system does not appear to be accurate despite troubleshooting (e.G.Resetting the system), do not rely on the navigation system.Reliance on data provided by a camera with an interrupted optical path may lead to inaccurate measurements.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE PULSE SYSTEM
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key15033074
Report Number2031966-2022-00091
Device Sequence Number6920356
Product Code OLO
UDI-Device Identifier00887517990792
UDI-Public887517990792
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K210574
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1881500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Initial Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/15/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-