• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THE SPECTRANETICS CORPORATION ANGIOSCULPT OTW PTA; CATHETER, PTA, CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THE SPECTRANETICS CORPORATION ANGIOSCULPT OTW PTA; CATHETER, PTA, CUTTING/SCORING Back to Search Results
Model Number 2092-4040
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
The angiosculpt device was used in a peripheral lesion.The balloon was inflated to 10 atm for 30 seconds with no issues.Then, the balloon was pressurized to 14 atm but the value on the indeflator did not go up due to a rupture.The procedure was completed with a non-angiosculpt device.No patient injury reported.This product problem is being submitted conservatively because the balloon ruptured at rbp.
 
Additional Manufacturer Narrative
Report source: foreign- japan.Visual inspection found no unusual characteristics of the device.During functional testing, using an indeflator filled with green color dye water, the balloon was inflated to 10 atm and a pinhole leak was confirmed on the proximal portion of the balloon.The ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGIOSCULPT OTW PTA
Common Device Name
CATHETER, PTA, CUTTING/SCORING
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
5055 brandin court
fremont CA 94538
Manufacturer (Section G)
SPECTRANETICS
5055 brandin court
fremont CA 94538
Manufacturer Contact
ana tan
5055 brandin court
fremont, CA 94538
MDR Report Key15033608
Report Number3005462046-2022-00030
Device Sequence Number6929213
Product Code PNO
UDI-Device Identifier00813132021405
UDI-Public00813132021405
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100303
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/15/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/12/2025
Device Model Number2092-4040
Device Catalogue Number2092-4040
Device Lot NumberG22010018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Initial Report FDA Received Date07/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/12/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK MFG AND SIZE: GUIDE CATHETER; UNK MFG AND SIZE: GUIDEWIRE; UNK MFG AND SIZE: INTRODUCER SHEATH
Patient SexUnknown
-
-