Report source: foreign- japan.Visual inspection found no unusual characteristics of the device.During functional testing, using an indeflator filled with green color dye water, the balloon was inflated to 10 atm and a pinhole leak was confirmed on the proximal portion of the balloon.The ifu warns at least 99.9% of the balloons (with a 95% confidence level) will not burst at or below the rbp.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|