| Model Number |
N/A |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/21/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of asftf116 showed no other similar product complaint(s) from this lot number.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer "patient presented to clinic with needle intact/port accessed.Dressing intact, able to flush port, blood return obtained.While at home, patient stated that he noticed that the iv tubing had become completely dislodged from the primary hub connection site on the infusion set.".
|
| |
|
Additional Manufacturer Narrative
|
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and batch history, applicable previous investigation(s), photo analysis and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a detached hub on the infusion set was confirmed but the cause is unknown.A photograph of the implicated 20g miniloc safety infusion set was returned for evaluation.The infusion set contained characteristics of use (e.G., residual material within the tubing and y-site, an activated safety mechanism, and auxiliary devices attached to the infusion set).The side tail luer adaptor had completely detached from the tubing.It could not be determined from the photograph if the tubing had broken near the joint or if the tubing had dislodged from the luer adaptor.The observable features on the submitted photograph did not contain enough information to identify a specific root cause of the reported event.Possible contributing factors could include tensile (pulling) damage, material fatigue, or damage from a sharp instrument.H3 other text : evaluation findings are in section h.11.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer "patient presented to clinic with needle intact/port accessed.Dressing intact, able to flush port, blood return obtained.While at home, patient stated that he noticed that the iv tubing had become completely dislodged from the primary hub connection site on the infusion set." additional information received: the tubing was ¿tied¿ off by patient when he realized the tubing became completely dislodged.Date of occurrence: (b)(6) 2022.Placement date: (b)(6) 2022 explant date: (b)(6) 2022 was there any patient harm reported? if yes, please describe the harm, how it was treated and the outcome.No harm; however, the potential for harm was 100%.
|
| |
|