Model Number 20028E
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd alaris smartsite extension set experienced occlusion.The following information was provided by the initial reporter: cannot flush through extension set.
 
Additional Manufacturer Narrative
Investigation summary one sample extension set of model # 20028e, lot 21119585 was received for quality investigation.The set was examined for defects and abnormalities.No defects or abnormalities were observed.The set was attached to a cut off bd 10 ml syringe to see if a fluid flow path can be observed and also observed that fluid is flowing clearly through filter with slight push to the syringe and the set was attempted to be flushed.The set was successfully flushed.The customer complaint that the extension set with flow issue -fluid blockage could not be replicated.A device history record review for model 20028e and lot number 21119585 was performed.The search showed that a total of (b)(4).There were no quality notifications issued for the failure mode reported by the customer during the production build of this set of lot.The root cause of flow issues - fluid blockage occurs could not be definitively determined.
 
Event or Problem Description
It was reported that the bd alaris smartsite extension set experienced occlusion.The following information was provided by the initial reporter: cannot flush through extension set.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15033945
Report Number9616066-2022-00902
Device Sequence Number6911921
Product Code FPA
UDI-Device Identifier07613203012393
UDI-Public07613203012393
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/25/2024
Device Model Number20028E
Device Catalogue Number20028E
Device Lot Number21119585
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/16/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date07/15/2022
Supplement Report FDA Received Date08/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-