• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 4/12 MM R; KNEE TIBIAL INSERT FIXED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 4/12 MM R; KNEE TIBIAL INSERT FIXED Back to Search Results
Model Number 02.12.0412FR
Health Effect - Clinical Code Joint Laxity (4526)
Date of Event 06/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery at about 1 year and 2 months post primary for knee instability.The surgeon explanted the liner thick 12mm and implanted a thicker liner of 17mm.
 
Additional Manufacturer Narrative
Batch review performed on 05 july 2022: lot 2002795: 75 items manufactured and released on 10-june-2020.Expiration date: 2025-06-01.No anomalies found related to the problem.To date, 62 items of the same lot have been sold without any similar reported event during the period of review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GMK-SPHERE TIBIAL INSERT FIXED SPHERE FLEX SIZE 4/12 MM R
Common Device Name
KNEE TIBIAL INSERT FIXED
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15037411
Report Number3005180920-2022-00553
Device Sequence Number17878278
Product Code JWH
UDI-Device Identifier07630030826580
UDI-Public07630030826580
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K121416
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.12.0412FR
Device Catalogue Number02.12.0412FR
Device Lot Number2002795
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Initial Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
Patient RaceWhite
-
-