"this report is part of a retrospective review and remediation efforts in response to a warning letter.Recall: z-0956-2020.(b)(4).S/w 3.18e.Device received with a depleted duracell alkaline battery installed.Device passed the functional test, including the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and the dat test at 0.08715 inches.Device had a corroded battery tube, scratched case, pillowing keypad overlay and a cracked retainer.Data analysis: (date of customer passing: 8/4/18) 07/29/2018 dailytotalofallinsulindelivered = 194.15.07/30/2018 dailytotalofallinsulindelivered = 159.425.07/31/2018 dailytotalofallinsulindelivered = 38.475.08/01/2018 dailytotalofallinsulindelivered = 73.725.08/02/2018 dailytotalofallinsulindelivered = 55.175.08/03/2018 dailytotalofallinsulindelivered = 0.08/04/2018 dailytotalofallinsulindelivered = 0.
|