MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYMMETRY; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYMMETRY; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 23295
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in a shunt in the non-tortuous and moderately calcified brachial vein.A 3.5-4/4t/90 symmetry balloon catheter was advanced for dilation.However, during first inflation at 18 atmospheres for 30 seconds, the balloon ruptured.The device was removed and the procedure was completed with different device.No patient complications were reported and the patient condition was good.
 
Additional Manufacturer Narrative
Initial reporter address 1: (b)(6).
 
Event or Problem Description
It was reported that balloon rupture occurred.The 90% stenosed target lesion was located in a shunt in the non-tortuous and moderately calcified brachial vein.A 3.5-4/4t/90 symmetry balloon catheter was advanced for dilation.However, during first inflation at 18 atmospheres for 30 seconds, the balloon ruptured.The device was removed and the procedure was completed with different device.No patient complications were reported and the patient condition was good.
 
Additional Manufacturer Narrative
E1 initial reporter address (b)(6).Device evaluated by mfr.: a symmetry device was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.A longitudinal tear was identified in the balloon material.The tear measured 6mm in length and extended from a position 3mm distal of the proximal markerband to 2mm proximal of the proximal markerband.The rated burst pressure for this balloon is 15 atmospheres as per symmetry specification.A visual and tactile examination identified no damage or issues with the shaft of the device.A visual examination identified no issues with the tip of the device that could have contributed to the complaint incident.This concludes the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMMETRY
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15040106
Report Number2124215-2022-26036
Device Sequence Number6943899
Product Code LIT
UDI-Device Identifier08714729181736
UDI-Public08714729181736
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K143193
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/17/2025
Device Model Number23295
Device Catalogue Number23295
Device Lot Number0029421126
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer08/05/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date08/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-