| Model Number |
03.27.139 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
At about 1 month and 3 weeks after the primary surgery, the patient came in for a post-op appointment and it was observed that a cage was backing out of disc space.The surgeon removed the cage, caps, and rod and implanted a competitor expandable cage.The existing rod was re-inserted and new caps were used to lock it down.The surgery was completed successfully.The patient did not follow post-op guidelines and the patient's lifestyle contributed to the event.
|
| |
|
Additional Manufacturer Narrative
|
|
Batch review performed on 30-jun-2022: lot 1822376: (b)(4) items manufactured and released on 17-jun-2019.Expiration date: 2024-06-01.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.
|
| |
|