Event or Problem Description
|
|
It was reported that on (b)(6)2022, while inspecting the instruments after going through the decontamination process, there were several devices damaged.The t-handle chuck and the handle bent.The power shaft screwdriver and the tip stripped.There was a long reamer shaft and the reamer heads will not fit/attach and the the 2.5mm reaming rod keeps getting caught in it.The cable tensioning attachment will not stay on the tensioner.There was also a locking small frag tray insert.One of the handles on the tray is broken.There is no patient or surgical involvement.This is report 3 of 3 for (b)(4).This report is for a tensioning device.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Event date is unknown; it is unknown when devices became damaged/worn as this was not an intraoperative event.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Initial reporter is j&j company representative the investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a product investigation was conducted.Note: the investigation was performed by the supplier.Although distributed by depuy synthes syn, the product is designed, manufactured, and labeled by the manufacturing supplier.The product was returned to depuy synthes for evaluation.The depuy synthes team forwarded the device to the supplier which conducted a visual inspection of the returned device.Upon initial inspection it was discovered that the device shows signs of use.When looking into the cam-lock body, that the canted coil spring is very deformed and will no longer allow the tensioning device to be attached to the mating tensioner properly.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the 391.884, provisional tensioning device.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.A device history record (dhr) review was conducted: part #: 391.884 synthes lot #: p302644 supplier lot #: p302644 release to warehouse date: 17 may 2004 supplier: pioneer surgical technology no ncr's generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|