Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Device report from synthes reports an event in japan as follows: it was reported that on (b)(6) 2022, the patient underwent a surgery.When the surgeon was drilling a screw hole at most proximal of the distal screw holes of a mhn long nail with the drill bit in question, the drill bit interfered with the nail and the tip of the drill bit broke.The broken tip of the drill bit was removed from the patient¿s body successfully.This was confirmed by x-ray.The surgery was completed successfully with no surgical delay.Concomitant device reported: unk - nails (part# unknown; lot# unknown).This is report 1 of 1 for (b)(4) this report is for a drill bit.
|