• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC FLEXOR RAABE GUIDING SHEATH; DYB INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC FLEXOR RAABE GUIDING SHEATH; DYB INTRODUCER, CATHETER Back to Search Results
Model Number G12266
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
As reported, prior to a lower limb angioplasty, a flexor raabe guiding sheath's dilator tip was damaged/split.This was verified on the instrumentation table and the device did not make patient contact.The device was replaced with a new same-type device, without causing any harm to the patient.A photo provided by the customer shows that the tip of the dilator appears to be split.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects.
 
Additional Manufacturer Narrative
Name and address - postal code: (b)(6), phone: (b)(6).This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary: as reported, prior to a lower limb angioplasty, a flexor raabe guiding sheath's dilator tip was damaged/split.This was verified on the instrumentation table and the device did not make patient contact.The device was replaced with a new same-type device, without causing any harm to the patient.A photo provided by the customer shows that the tip of the dilator appears to be split.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects.Investigation - evaluation: reviews of the complaint history, device history record, instructions for use, manufacturing instructions, and quality control procedures of the device were conducted during the investigation.No device was returned for investigation.A photo provided by the user confirms the dilator has a split in the distal tip.A document-based investigation evaluation was performed.One relevant nonconformance was recorded for this product lot; the lone affected device was scrapped.There have been no other reported complaints for this lot number.The device history record review provides objective evidence that the device was manufactured to specification.There is no evidence of nonconforming devices from the complaint lot in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Sufficient inspection activities are in place to identify this failure mode prior to distribution.The device is provided with instructions for use which state, ¿upon removal from package, inspect the product to ensure no damage has occurred.¿ based on the available information, cook has concluded that a component failure contributed to this failure mode.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.Per a review of the risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXOR RAABE GUIDING SHEATH
Common Device Name
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15042451
Report Number1820334-2022-01223
Device Sequence Number15197809
Product Code DYB
UDI-Device Identifier00827002122666
UDI-Public(01)00827002122666(17)240908(10)14199298
Combination Product (Y/N)N
PMA/510(K) Number
K142829
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/08/2024
Device Model NumberG12266
Device Catalogue NumberKCFW-6.0-38-90-RB-RAABE
Device Lot Number14199298
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer10/17/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date11/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-