MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC PROPHECY INBONE PATIENT SPECIFIC GUIDES; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC PROPHECY INBONE PATIENT SPECIFIC GUIDES; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number PROPINB
Medical Device Problem Code Inaccurate Information (4051)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 06/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis results are not available at the time of this report.A follow-up report will be sent when the analysis is complete.
 
Event or Problem Description
It was reported that the label on the alignment guide did not match the patient case number.The guide was labeled for another case.
 
Additional Manufacturer Narrative
Corrections: d9/h9, h3.The reported event could be confirmed, since intraoperative images provided do show the wrong case number on the tibial guide.Information provided by the rep indicates the surgeries were performed 3 weeks apart, which indicates mix-up at the hospital is unlikely.Based on investigation, the root cause was attributed to a product mix-up related issue.The failure was caused by the parts being mixed during production inspection or packaging.It should be noted, all prophecy cases are accompanied by standard surgical instrumentation for use by the surgeon if needed.A nonconformance investigation was initiated to investigate in further detail and determine any additional necessary actions.
 
Event or Problem Description
It was reported that the label on the alignment guide did not match the patient case number.The guide was labeled for another case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROPHECY INBONE PATIENT SPECIFIC GUIDES
Common Device Name
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15042733
Report Number3010667733-2022-00249
Device Sequence Number15185818
Product Code HSN
UDI-Device Identifier00889797057110
UDI-Public00889797057110
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K110360
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/22/2022
Device Model NumberPROPINB
Device Catalogue NumberPROPINB
Device Lot Number1731247
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2022
Supplement Date Received by Manufacturer09/13/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date10/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-