AIZU OLYMPUS CO., LTD. FORCEPS/IRRIGATION PLUG (ISOLATED TYPE); FORCEPS/ IRRIGATION PLUG
|
Back to Search Results |
|
Model Number MAJ-891 |
Device Problem
Failure to Clean Adequately (4048)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/15/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer was provided with the correct reprocessing material for the device.Customer was also advised that the device could be discarded and a maj-2092 be used instead.The device is not returned, as such a definitive root cause of the reported complaint cannot be determined at this time.This event is under investigation.A supplemental report will be submitted upon completion of the investigation or upon receiving additional information.
|
|
Event Description
|
As reported for this event, during a hygiene audit at the facility, it was observed that the device was not being disassembled at the time of reprocessing.Dirt was observed inside the device.There is no harm to any patient.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record (dhr) was unable to be reviewed for this device however, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, the user¿s understanding differed from olympus recommendation in device handling and reprocessing steps.The following is included in the instructions for use (ifu): "be sure to disassemble the forceps/irrigation plug before it is cleaned, disinfected or sterilized.Otherwise, the forceps/irrigation plug may not be properly cleaned/disinfected/sterilized".Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|