• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WAVELIGHT GMBH WAVELIGHT FS200 EASYPACK PATIENT INTERFACE; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WAVELIGHT GMBH WAVELIGHT FS200 EASYPACK PATIENT INTERFACE; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 1515
Medical Device Problem Code Decrease in Suction (1146)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 04/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
A facility representative a patient with loss of suction with one interface.There are multiple related reports for this facility.This report addresses a pi (b)(4) and additional manufacturer reports will be filed.
 
Additional Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The review of logfile for the day of treatment shows all laser system functions were within specifications.The vacuum check and the energy check were performed successfully without issue.The ablation test was performed successfully and without problems, but the ablation test image is blacked out because the microscope illumination was not turned on, therefore the ablation test image does not show any steps between the orientation lines.The logfile shows that the user performed one energy check at system startup and then performed fourteen treatments without further energy check on that day.The reported treatment could be identified in the logfile.The treatment was aborted by user after the bcc (beam control check).Suction process was not started.Malfunction cannot be confirmed.No technical root cause could be identified.Review of the logfiles for the treatment day shows no relevant warning or error messages.The system was working within specification.The user restarted the treatment on the same day with a new patient interface and finished the treatment without any problems.No sample evaluation is required even if the sample is returned as the root cause has been identified during logfile review.No technical root cause was identified as the product was found to be within specifications.The root cause is user handling.The treatment was stopped by user and could be finished successfully using another patient interface.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT FS200 EASYPACK PATIENT INTERFACE
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15042923
Report Number3003288808-2022-00202
Device Sequence Number5139312
Product Code GEX
UDI-Device Identifier00380659982989
UDI-Public00380659982989
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K190392
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1515
Device Catalogue Number8065998299
Device Lot Number144N2K
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date11/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-