Model Number G14521
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during an unknown procedure, an open-end flexi-tip ureteral catheter broke during use.A ureteroscopy had to be performed to retrieve the broken piece successfully.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Initial reporter: occupation: inventory coordinator.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Additional Manufacturer Narrative
Event summary: it was reported that during an unknown procedure, an open-end flexi-tip ureteral catheter broke during use.A ureteroscopy had to be performed to retrieve the broken piece successfully.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Investigation - evaluation: a document-based investigation was performed including a review of complaint history, device history record (dhr), quality control data, and the instructions for use (ifu).The complaint device was not returned; therefore, no physical examinations could be performed.A review of the device history record found no non-conformances related to the reported failure mode.Because there are no related non-conformances, adequate inspection activities have been established, there is objective evidence that the dhr was fully executed, and no other related complaints from the lot have been received from the field, it was concluded that there is no evidence that nonconforming product exists in house or in field.A review of complaint history records shows no other complaints associated with the complaint device lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: precautions: avoid bending or kinking the catheter prior to placement.Doing so could damage the integrity of the catheter and result in patient injury.If you encounter resistance while advancing or withdrawing the catheter, stop.Determine the cause of the resistance before proceeding.Do not withdraw catheter while deflected in cystoscope/endoscope.A review of the device history records found no anomalies and no other complaint have been reported for the lot.A review of manufacturing procedures found controls to be in place to assure functionality and device integrity prior to shipping.Based on the limited information surrounding the event a cause for the complaint could not be established it was not possible to rule out inadvertent user/procedural issues.Per the quality engineering risk assessment, no further action is warranted.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPEN-END FLEXI-TIP URETERAL CATHETER
Common Device Name
KOD CATHETER, UROLOGICAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15043386
Report Number1820334-2022-01224
Device Sequence Number6929690
Product Code KOD
UDI-Device Identifier00827002145214
UDI-Public(01)00827002145214(17)250420(10)14680146
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K171662
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/20/2025
Device Model NumberG14521
Device Catalogue Number021305
Device Lot Number14680146
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/13/2022
Supplement Date Received by Manufacturer07/20/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date07/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexUnknown
-
-