| Model Number |
1804325-23 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Myocardial Infarction (1969); Sepsis (2067)
|
| Date of Event |
03/21/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient has a history of st elevated myocardial infarction (stemi) and on (b)(6) 2022 a 3.5x23 mm xience skypoint stent was implanted.On (b)(6) 2022 the patient was re-hospitalized due to an acute myocardial infarction and unspecified sepsis.Medication was provided and no additional patient effects were reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The stent remains in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Based on the information reviewed, there is no indication of a product quality issue.The reported patient effects of sepsis and myocardial infarction are listed in the xience skypoint, everolimus eluting coronary stent system (eecss), electronic instructions for use (eifu) as known patient effects of coronary stenting procedures.A conclusive cause for the reported patient effects, and the relationship to the product, if any, cannot be determined; however, the reported treatment appears to be related to the operational context of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient has a history of st elevated myocardial infarction (stemi) and on (b)(6) 2022 a 3.5x23 mm xience skypoint stent was implanted.On (b)(6) 2022 the patient was re-hospitalized due to an acute myocardial infarction and unspecified sepsis.Medication was provided and no additional patient effects were reported.Subsequent to the initially filed report, the following information was provided: the patient was discharged from the initial procedure on (b)(6) 2022 and was discharged from readmission on (b)(6) 2022.No additional treatment was performed other than the medication provided.No additional information was provided.
|
| |
|