| Model Number |
7121/65 |
| Medical Device Problem Codes |
Fracture (1260); Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
| Date of Event |
07/12/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During follow-up, noise leading to oversensing and the delivery of inappropriate high voltage therapy was observed on the right ventricular (rv) lead.Lead damage was suspected but was not confirmed visually.The rv lead was explanted and replaced to resolve the event.The patient was stable and there were no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported events of lead fracture and noise resulting in oversensing and inappropriate high voltage therapy were not confirmed.As received, a partial lead was returned in two pieces.Electrical testing did not find any indication of conductor fractures or internal shorts.
|
| |
|