U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS ULTRASOUND EPIQ CVXI; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS ULTRASOUND EPIQ CVXI; SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC Back to Search Results
Model Number EPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Medical Device Problem Code Application Program Freezes or Fails to Launch (4031)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is underway to determine the root cause of the issue.A follow up report will be provided upon investigation completion.
 
Event or Problem Description
A customer reported an epiq cvxi ultrasound system using an x8-2t transducer shut down and the display went black during a heart procedure.They were unable to reboot the system and another ultrasound system was brought in to complete the procedure using the same transducer.There was no patient or user harm associated with this event.
 
Additional Manufacturer Narrative
Failure analysis of the returned acquisition module was performed by the vendor and they could not find any faults with the part.However, replacement of the acquisition module resolved the immediate issue.No further issues reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ CVXI
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key15044438
Report Number3019216-2022-00069
Device Sequence Number5248479
Product Code IYN
UDI-Device Identifier00884838097940
UDI-Public00884838097940
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Device Catalogue Number795232
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/25/2022
Supplement Date Received by Manufacturer05/25/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date05/15/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/08/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-