| Model Number |
6500 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Operator of device is patient/consumer.
|
| |
|
Event or Problem Description
|
|
It was reported that the a high pressure alarm went off.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: event problem and evaluation codes: updated.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection found the device in good condition.Error was found in the device ehl (event history log) multiple times.The customer stated problem was not duplicated.No device problem was found.Preventively, replaced the downstream occlusion sensor and the upstream occlusion sensor was re-calibrated.The cause of the reported problem could not be determined.A dhr (device history review) was not conducted, based upon review of the information provided by the customer, as it does not indicate a problem with the initial manufacture or prior repair of the device., corrected data: d1: brand name; d2: common device name; d2: product code; d4: model number; g5: premarket (510k) number; h4: device manufacturing date.
|
| |
|