• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE TMZF HIP STEM #5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6020-0537
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Insufficient Information (4580)
Date of Event 04/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported through claimant's counsel that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2011 and was revised on (b)(6) 2022.During the revision the surgeon noted "¿ there was a significant amount of black metallosis fluid encountered.An extensive debridement of all metal debris was performed and it was found that his trunnion was completely destroyed.".
 
Event or Problem Description
It was reported through claimant's counsel that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2011 and was revised on (b)(6) 2022.During the revision the surgeon noted "¿ there was a significant amount of black metallosis fluid encountered.An extensive debridement of all metal debris was performed and it was found that his trunnion was completely destroyed.".
 
Additional Manufacturer Narrative
Reported event: an event regarding wear involving an accolade stem that was mated with an lfit v40 cocr head was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: a review of the device history records indicates the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: no other similar events were reported for the lot.Conclusion: the reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of an nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE TMZF HIP STEM #5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key15046515
Report Number0002249697-2022-01027
Device Sequence Number5646469
Product Code LPH
UDI-Device Identifier04546540510570
UDI-Public04546540510570
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2015
Device Model Number6020-0537
Device Catalogue Number6020-0537
Device Lot Number34732206
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer04/27/2023
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date05/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/24/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient SexMale
-
-