| Model Number |
24705 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 80% stenosed, target lesion was located in the moderately tortuous and moderately calcified anterior tibial artery.A 3.0mm x 150mm x 150cm coyote balloon catheter was advanced for dilatation.However, during the first inflation at 10 atmospheres, the balloon ruptured.The procedure was completed with a 3x150 coyote balloon catheter.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a coyote balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed no damages.Microscopic examination revealed a pinhole 12.4cm from the tip.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 80% stenosed, target lesion was located in the moderately tortuous and moderately calcified anterior tibial artery.A 3.0mm x 150mm x 150cm coyote balloon catheter was advanced for dilatation.However, during the first inflation at 10 atmospheres, the balloon ruptured.The procedure was completed with a 3x150 coyote balloon catheter.No patient complications were reported.
|
| |
|