BOSTON SCIENTIFIC CORPORATION MACH1; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 1260
Medical Device Problem Codes Fracture (1260); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/24/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a tip fracture occurred.While unpacking a 6f mach 1 cls3.5 guide catheter, a tip fracture occurred.No patient complications were reported in relation to this event.
 
Additional Manufacturer Narrative
Device evaluated by manufacturer: visual and microscopic observation was performed on device.There were numerous kinks on shaft; however, tip was observed to have no fracture.No other abnormalities were observed.The reported events cannot be confirmed as tip was observed to be intact and with no fracture.
 
Event or Problem Description
It was reported that a tip fracture occurred.While unpacking a 6f mach 1 cls3.5 guide catheter, a tip fracture occurred.No patient complications were reported in relation to this event.It was further reported that the catheter tip was broken while it was unpacked and was not used on the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MACH1
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
FLEXTRONICS INTERNATIONL EUROPE B V
6201 america center drive
san jose CA 95002
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15047831
Report Number2134265-2022-07976
Device Sequence Number12090661
Product Code DQY
UDI-Device Identifier08714729350644
UDI-Public08714729350644
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
K020028
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2024
Device Model Number1260
Device Catalogue Number1260
Device Lot Number0060323622
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/28/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date08/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-