SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD SMARTSITE¿ EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 20159E
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/02/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the bd smartsite¿ extension set experienced blockage.The following information was provided by the initial reporter: i reached out to customer service about item #20159e that was not flushing.There are two lot numbers that clinical staff has identified as having this issue with multiple items.
 
Event or Problem Description
It was reported that the bd smartsite¿ extension set experienced blockage.The following information was provided by the initial reporter: i reached out to customer service about item #20159e that was not flushing.There are two lot numbers that clinical staff has identified as having this issue with multiple items.
 
Additional Manufacturer Narrative
H6: investigation summary three samples (model #20159e) were returned by the customer.It was reported that the set was not flushing.The sets were examined for defects and abnormalities.No defects or abnormalities were observed.The samples were connected to a cut off syringe tip to check that the fluid path is in an activated position/open when connected to a syringe.The samples were then attempted to be flushed with water using a 10ml bd syringe to confirm an open fluid path.The fluid paths of all samples were open and those samples were able to be flushed.The failure was unable to be replicated, and the complaint could not be verified.A device history record review for model 20159e lot number 21119807 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 08nov2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A root cause was unable to be determined because the failure was unable to be replicated.Due to an increase in incidents of this failure mode, a trend for this occlusion issue has been identified for this product line, a capa (corrective action preventative action) 1998036 has been initiated, and a team has been assembled in order to investigate the issue.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD SMARTSITE¿ EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15047879
Report Number9616066-2022-00910
Device Sequence Number11967388
Product Code FPA
UDI-Device Identifier37613203012288
UDI-Public37613203012288
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date11/05/2024
Device Model Number20159E
Device Catalogue Number20159E
Device Lot Number21119806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date10/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-